MOSAIQ <strong>Dosimetrist</strong>/<strong>Physicist</strong> <strong>Training</strong> <strong>Guide</strong> Module I<br />

Notes<br />

Figure 85: Diagnoses and Interventions Window with Two Treatment Fields<br />

Manually Adding a Treatment Field<br />

In most cases, you will copy simulation fields or other treatment fields to add<br />

treatment fields. You can add a treatment field “from scratch” if the patient treatment<br />

does not come over from your treatment planning system through DICOM. This<br />

could occur if the patient was not planned out, or if you do not transfer plans from a<br />

treatment planning system into MOSAIQ. In the following exercises, you will<br />

manually add a treatment field and manually add geometric and other information.<br />

Manually Adding Treatment Fields<br />

In this exercise, you will manually add a treatment field.<br />

1. In the Diagnoses and Interventions window, click the Lung Boost radiation<br />

prescription.<br />

Specifies the radiation prescription to which to manually add the field.<br />

2. Click .<br />

Opens the Treatment Field Definition dialog box.<br />

3. Type 3 in the first Field text box, then press TAB.<br />

Specifies the Treatment Field ID, which identifies the treatment field.<br />

4. Type LAO LUNG in the second Field text box.<br />

Specifies a more detailed description of the treatment field.<br />

5. Click adjacent to the Machine field, then double-click a linear accelerator.<br />

Specifies the linear accelerator used to treat the patient. When you select the linear<br />

accelerator, you can also click the BEV radio button to open the Beam’s Eye<br />

View. It shows the dimensions of the treatment field.<br />

102 Copyright 2009, IMPAC Medical Systems<br />

LTGRAD0005A/2.0/26-Feb-2009