MOSAIQ <strong>Dosimetrist</strong>/<strong>Physicist</strong> <strong>Training</strong> <strong>Guide</strong> Module I<br />

Notes<br />

Changing Approved Treatment Fields<br />

If the machine characterization of the treatment machine or patient treatment plan<br />

changes, you may have to change an approved treatment field definition. If you<br />

change an approved field, the status of the field automatically changes to Unapproved.<br />

You have to re-approve the field before treatment, unless you configure MOSAIQ to<br />

let you copy couch parameters and change field notes without changing the status.<br />

Both are included in the exercise that follows. When you have re-approved the<br />

treatment field, you can open the Field Delta View dialog box to show all versions of<br />

the treatment field. In this exercise, you will change an approved treatment field, reapprove<br />

it, then open the Field Delta View dialog box.<br />

1. In the Diagnoses and Interventions window, double-click the 3 LAO Lung<br />

treatment field.<br />

Opens the Treatment Field Definition dialog box.<br />

2. Type 90 in the Collimator Angle field.<br />

Specifies the collimator angle for the treatment field.<br />

3. Type 3 in the Open MU field.<br />

Specifies the MU for the open portal image.<br />

4. Type NONE in the Bolus field.<br />

Specifies that no bolus for this treatment field, shown in Figure 95.<br />

Figure 95: Approved Treatment Field with Changes<br />

112 Copyright 2009, IMPAC Medical Systems<br />

LTGRAD0005A/2.0/26-Feb-2009