MOSAIQ <strong>Dosimetrist</strong>/<strong>Physicist</strong> <strong>Training</strong> <strong>Guide</strong> Module I<br />
Notes 12. Type open field portal image monitor units in the Open MU field.<br />
13. Type the portal image delta setting in the Open Delta field.<br />
Adding Accessory and Geometric Settings<br />
1. In the Accessories/Slots group of the Treatment Field Definition dialog box, click<br />
the type of wedge from the Wedge field.<br />
2. Type the compensator in the Compensator field.<br />
3. Type the block needed for treatment in the Block field.<br />
4. Type the bolus in the Bolus field.<br />
5. Change any geometric settings if necessary.<br />
Applying a Tolerance Table<br />
1. In the Treatment Field Definition dialog box, click adjacent to the Tolerance<br />
field to open the Tolerance Table list.<br />
2. Click the tolerance table for the treatment field.<br />
Adding Treatment Field Notes<br />
1. In the Treatment Field Definition dialog box, click to show the Field<br />
Setup view.<br />
2. Click in the Field Setup text box, then type the field note content.<br />
3. Click to save the note and the treatment field.<br />
Using the Apply Limits Option<br />
1. Double-click the treatment field to open the Treatment Field Definition dialog<br />
box.<br />
2. Change one or two geometric settings so that they are out of bounds.<br />
3. Right-click or press F4 to open the Treatment Field Definition Options menu.<br />
4. Click Apply Limits….<br />
5. Click if the Positional Limit Exceeded message box appears.<br />
124 Copyright 2009, IMPAC Medical Systems<br />