1/1/2005 Chapter 8 Corneal Curvature and Topography page 8-23<br />

The Mastervue unit is an alternative means <strong>of</strong> collecting videokeratography data<br />

for clinics not equipped with a TMS unit. It is currently being utilized at the Nova<br />

Southeastern Participating Clinic. Data are stored on site.<br />

8.7.1 Calibration Verification<br />

Calibration <strong>of</strong> the Mastervue unit is performed weekly throughout the course <strong>of</strong><br />

data collection for the CLEK Study.<br />

The calibration procedure is as follows:<br />

(1) Place the frame <strong>of</strong> the calibration bar in the two holes located at the base <strong>of</strong><br />

the chin rest.<br />

(2) Select Utilities from the Main Menu.<br />

(3) Select Run Calibration from the Utilities Submenu.<br />

(4) Position the projection head in front <strong>of</strong> the 35 D sphere.<br />

(5) Select OK.<br />

(6) Align the projection head by centering the cross hairs in the smallest ring <strong>of</strong><br />

the image. Use the joystick to focus the image. Press the button on the joystick to<br />

capture the image. If the image capture is within an acceptable value, the word<br />

“passed” will be displayed on the screen; if not within an acceptable value, the word<br />

“over” or “under” will be displayed along with the actual dioptric value. To continue<br />

the calibration procedure, select Yes; to repeat the measurement <strong>of</strong> the 35 D sphere,<br />

select No.<br />

(7) Repeat the above procedure for the 42.25 D and 55 D spheres.<br />

(8) Final results will be displayed on the screen. Maintain a record documenting<br />

weekly calibration. If the machine is improperly calibrated, contact Humphrey<br />

Instruments Technical Support for assistance at 800-227-1508. If machine cannot be<br />

recalibrated properly prior to patient visit, continue with the examination protocol.<br />

Following the examination, complete a Feedback Report documenting the calibration<br />

results only if a calibration error occurred. Submit the Feedback Report with patient<br />

record and maintain copies in the Clinic patient record and Clinic calibration record.<br />

8.7.2 Humphrey Mastervue Patient Confidentiality<br />

Patient confidentiality is maintained as follows: