1/1/2005 Chapter 8 Corneal Curvature and Topography page 8-23<br />
The Mastervue unit is an alternative means <strong>of</strong> collecting videokeratography data<br />
for clinics not equipped with a TMS unit. It is currently being utilized at the Nova<br />
Southeastern Participating Clinic. Data are stored on site.<br />
8.7.1 Calibration Verification<br />
Calibration <strong>of</strong> the Mastervue unit is performed weekly throughout the course <strong>of</strong><br />
data collection for the CLEK Study.<br />
The calibration procedure is as follows:<br />
(1) Place the frame <strong>of</strong> the calibration bar in the two holes located at the base <strong>of</strong><br />
the chin rest.<br />
(2) Select Utilities from the Main Menu.<br />
(3) Select Run Calibration from the Utilities Submenu.<br />
(4) Position the projection head in front <strong>of</strong> the 35 D sphere.<br />
(5) Select OK.<br />
(6) Align the projection head by centering the cross hairs in the smallest ring <strong>of</strong><br />
the image. Use the joystick to focus the image. Press the button on the joystick to<br />
capture the image. If the image capture is within an acceptable value, the word<br />
“passed” will be displayed on the screen; if not within an acceptable value, the word<br />
“over” or “under” will be displayed along with the actual dioptric value. To continue<br />
the calibration procedure, select Yes; to repeat the measurement <strong>of</strong> the 35 D sphere,<br />
select No.<br />
(7) Repeat the above procedure for the 42.25 D and 55 D spheres.<br />
(8) Final results will be displayed on the screen. Maintain a record documenting<br />
weekly calibration. If the machine is improperly calibrated, contact Humphrey<br />
Instruments Technical Support for assistance at 800-227-1508. If machine cannot be<br />
recalibrated properly prior to patient visit, continue with the examination protocol.<br />
Following the examination, complete a Feedback Report documenting the calibration<br />
results only if a calibration error occurred. Submit the Feedback Report with patient<br />
record and maintain copies in the Clinic patient record and Clinic calibration record.<br />