2/22/96 Chapter 9 Contact Lens Fit Assessment page 9-5<br />

Table 9-2. Tolerances <strong>of</strong> fabrication for CLEK Study Trial Lens Set<br />

Parameter<br />

Tolerance<br />

Base curve radius<br />

0.05 mm without distortion or<br />

warpage<br />

Power<br />

0.25 D<br />

Overall diameter<br />

0.1 mm<br />

Optic zone diameter<br />

0.1 mm<br />

<strong>Center</strong> thickness<br />

0.02 mm<br />

Surface quality<br />

No surface quality degradation is accept<br />

9.3 Verification <strong>of</strong> the CLEK Trial Contact Lens Set<br />

It is important that CLEK Study trial lenses be verified throughout the Study to<br />

confirm parameter stability. It is also imperative that after every use, each lens is<br />

returned to its appropriate storage case. To monitor parameter stability and lens<br />

labeling, all trial lens base curves must be verified monthly for base curve radius,<br />

warpage, and correct storage. The CLEK Trial Contact Lens Verification Form is used to<br />

record these findings at each Participating Clinic. It is completed monthly and stored<br />

on-site.<br />

The base curve radius can be verified using a radiuscope, keratometer, or<br />

handheld keratometer. Trial lens base curves should not be warped by more than 0.02<br />

mm and should not be more than 0.05 mm from the targeted or labeled value (Table 9-<br />

2). If the base curve findings fall outside these targeted parameters, contact the<br />

Chairman’s <strong>of</strong>fice for a replacement lens.<br />

The following protocol is used for verifying the order and parameter stability <strong>of</strong><br />

the CLEK Study Trial Lens Set on a monthly basis by Clinic personnel certified for rigid<br />

contact lens verification.<br />

(1) Use a radiuscope, keratometer, or handheld keratometer to analyze the base<br />

curve radius <strong>of</strong> each lens in the CLEK Study Trial Lens Set every month. The<br />

measurement <strong>of</strong> other lens parameters is not required.<br />

(2) Record the base curve radius in millimeters in the column labeled Analyzed<br />

Base Curve Radius opposite the target or labeled value on the CLEK Trial<br />

Contact Lens Verification Form.<br />

(3) Any lens found in the wrong case should be transferred to the appropriate<br />

case.<br />

(4) Any lens analyzed as at least 0.06 mm different from the target measurement<br />

on the lens case should be replaced. Report the lens to be replaced to the<br />

Chairman’s Office.