Surgivet® V1030 Hand-Held Pulse Oximeter
Surgivet® V1030 Hand-Held Pulse Oximeter
Surgivet® V1030 Hand-Held Pulse Oximeter
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Chapter 1: Introduction<br />
Notes<br />
NOTE: The WW1090 rechargeable battery pack utilizes circuitry that optimizes the<br />
charging of the batteries. New packs will require multiple charge / discharge<br />
learning cycles before optimum performance is obtained.<br />
NOTE: Alarms may be tested while the monitor is in use by setting alarm limits such that<br />
the measured parameter is outside alarm limits. Return limits to the required<br />
settings after testing.<br />
NOTE: “SpO 2 averaging” means the number of pulse beats over which the SpO 2 value is<br />
averaged; “pulse averaging” means the number of seconds over which the pulse<br />
value is averaged.<br />
NOTE: Increasing or decreasing the averaging setting has no effect on the data update<br />
rate.<br />
NOTE: Alarm limits are retained through power cycles, with the exception of the following<br />
note.<br />
NOTE: If the low SpO 2 limit is set to less than 90% it will be reset to 90% when the monitor<br />
is next powered on. The high SpO 2 limit will be adjusted to 91% if it is 90% or less.<br />
NOTE: If the alarm volume is set to a value less then 8 at power down, it will be reset to 8<br />
when the monitor is next powered on.<br />
NOTE: A Patient Simulator does not calibrate the monitor. The monitor does not require<br />
calibration. A Patient Simulator provides a known SpO 2 and pulse rate to the<br />
monitor that allows the monitor’s performance to be checked.<br />
NOTE: A Patient Simulator cannot be used to assess the accuracy of a pulse oximeter<br />
and/or sensor.<br />
1-8 Veterinary <strong>V1030</strong> <strong>Hand</strong>-<strong>Held</strong> <strong>Pulse</strong> <strong>Oximeter</strong> Operation Manual