Chapter 13: Specifications<br />

SpO 2<br />

SpO 2 Range: 0 - 100% Functional Saturation (Display Maximum: 99%)<br />

SpO 2 Resolution: 1 count (%SpO 2 )<br />

SpO 2 Averaging: 4, 8, or 16 pulses (default = 8)<br />

SpO 2 Alarm limits:<br />

Calibration:<br />

Sensors:<br />

SpO 2 Accuracy:<br />

High: off and 99-51, 1 step (%SpO 2 ) increments.<br />

Low: 50-99, 1 step (%SpO 2 ) increments.<br />

Factory calibrated over range 70% to 100% SpO 2<br />

using human blood<br />

samples to functional saturation. Test methods are available upon request.<br />

No in-service calibration is required.<br />

Red: 660 nm, 2 mW (typical)<br />

Infrared: 905 nm, 2-2.4 mW (typical)<br />

70 to 100% SpO 2 , less than 70% SpO 2 is undefined.<br />

Sensor 6<br />

SpO 2 Accuracy in<br />

Normal Conditions 1<br />

SpO 2 Accuracy in<br />

Low Perfusion Conditions 1, 3<br />

BCI® 3044 ±2 2 ARMS ±2 2 ARMS<br />

SurgiVet® V1703 ±2.5 4 ARMS ±2.5 4 ARMS<br />

SurgiVet® V1707 ±2 4 ARMS ±2 4 ARMS<br />

SurgiVet® V3078 ±3.6 2, 5 ARMS ±3.6 2, 5 ARMS<br />

1<br />

Because pulse oximeter measurements are statistically distributed, only about two-thirds of pulse oximeter equipment<br />

measurements can be expected to fall within the ± A RMS of the value measured by the CO-oximeter.<br />

2<br />

The oximeter has been validated on 10 human adult volunteers that did not have health problems and were nonsmokers.<br />

The study was conducted at oxygen concentrations evenly distributed over an SpO 2 range of 70-100%.<br />

3<br />

<strong>Pulse</strong> Amplitude 1% to 0.1%. Tested using industry standard simulator.<br />

4<br />

Accuracy based on testing using industry standard simulator.<br />

5<br />

Accuracy based on equivalent BCI® 3078 sensor.<br />

6<br />

Accuracy of the V1700 reflectance sensor is unspecified.<br />

<strong>Pulse</strong> Rate<br />

<strong>Pulse</strong> Rate Range:<br />

<strong>Pulse</strong> Rate Resolution:<br />

20-350 bpm<br />

1 bpm<br />

<strong>Pulse</strong> Rate Averaging: 8 or 16 seconds (default = 8)<br />

<strong>Pulse</strong> Rate Alarm limits:<br />

High 350-21 bpm in 1 bpm increments<br />

Low 20-350 bpm in 1 bpm increments<br />

Sensor 5<br />

PR Accuracy in Normal<br />

Conditions 1<br />

PR Accuracy in<br />

Low Perfusion Conditions 1<br />

SurgiVet® V1703, V1707 ±2 ARMS ±2 2 ARMS<br />

SurgiVet® V3078 3 ±2 ARMS ±3 4 ARMS<br />

1<br />

Because pulse oximeter measurements are statistically distributed, only about two-thirds of pulse oximeter equipment<br />

measurements can be expected to fall within the ± A RMS of the value measured by the CO-oximeter.<br />

2<br />

<strong>Pulse</strong> Amplitude 1% to 0.1%. Tested using industry standard simulator.<br />

3<br />

Accuracy based on equivalent BCI® 3078 sensor.<br />

4<br />

<strong>Pulse</strong> Amplitude 1% to 0.3%. Tested using industry standard simulator.<br />

5<br />

Accuracy of the V1700 reflectance sensor is unspecified.<br />

13-2 Veterinary <strong>V1030</strong> <strong>Hand</strong>-<strong>Held</strong> <strong>Pulse</strong> <strong>Oximeter</strong> Operation Manual