Surgivet® V1030 Hand-Held Pulse Oximeter
Surgivet® V1030 Hand-Held Pulse Oximeter
Surgivet® V1030 Hand-Held Pulse Oximeter
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>Pulse</strong> Amplitude Index<br />
Range:<br />
Display:<br />
Display Update Rate:<br />
Chapter 13: Specifications<br />
0.03% to 20.00% 1 (0.01% increments)<br />
Normal Sensitivity Mode<br />
<strong>Pulse</strong> amplitude Index is defined as PI = (100 × AC)/DC where AC<br />
is the alternating current (pulsatile component of the signal) and<br />
DC is direct current (non-pulsatile component of the signal). It is<br />
a relative measure of pulse-signal strength over time at a pulse<br />
oximeter monitoring site, and is non-pulsatile in nature.<br />
High sensitivity mode – oximeter operation is extended below the<br />
PI threshold of 0.03%.<br />
9-segment green / yellow LED bargraph, logarithmic scale<br />
10 Hz<br />
In Normal Sensitivity mode, the oximeter will display dashes (Invalid data) for SpO 2 and PR<br />
when the PI value is below 0.03%.<br />
In the High Sensitivity mode, the oximeter will continue to process and display SpO 2 and PR<br />
values for PI values below 0.03%. SpO 2 accuracy and PR accuracy are not specified for PI values<br />
below 0.03%. Low perfusion performance is verified by simulator testing.<br />
1<br />
As tested with Industry Standard Simulator.<br />
The PI value maps to a 9-segment bar graph as shown below. The two lowest bars (1&2) are bicolor<br />
(Yellow & Green), and bars 3 through 9 are Green.<br />
PI From<br />
<strong>Oximeter</strong><br />
Segments<br />
Lit<br />
0.00-<br />
0.07<br />
0.08-<br />
0.15<br />
0.16-<br />
0.31<br />
0.32-<br />
0.63<br />
0.64-<br />
1.27<br />
1.28-<br />
2.55<br />
2.56-<br />
5.11<br />
5.12-<br />
10.23<br />
10.24-<br />
20.47<br />
1Y 2Y 3G 4G 5G 6G 7G 8G 9G<br />
Printer<br />
Paper:<br />
Modes:<br />
Paper Speed:<br />
Trend Graphs:<br />
58mm, Thermal<br />
Real Time, Tabular Trend, Graphic Trend<br />
12.5 mm/sec<br />
Headers print with graphs<br />
Serial Data Output<br />
Sensor Connector<br />
Data Type:<br />
Serial Data Protocol:<br />
ASCII comma delimited string output at 1 Hz. Lines terminated with a<br />
carriage return.<br />
RS232C, 9600 baud, 1 start bit, 8 data bits, 1 stop bit, no parity.<br />
Veterinary <strong>V1030</strong> <strong>Hand</strong>-<strong>Held</strong> <strong>Pulse</strong> <strong>Oximeter</strong> Operation Manual 13-3