Chapter 11: Troubleshooting<br />
Power<br />
PROBLEM POSSIBLE CAUSE CORRECTIVE ACTION<br />
The oximeter doesn’t<br />
turn on.<br />
Batteries weak and/ or no AC<br />
power.<br />
Replace or recharge the batteries.<br />
Connect to source of AC power.<br />
The oximeter turns<br />
on, but display reads:<br />
L0<br />
Bat<br />
The oximeter turns<br />
off unexpectedly.<br />
The External Power<br />
indicator X is not<br />
lit (AC Power).<br />
The monitor operates<br />
on AC power, but not<br />
on battery power.<br />
The monitor displays<br />
27<br />
ERR when powered<br />
up.<br />
Battery run time is<br />
excessively short on a<br />
fully charged battery.<br />
Batteries not installed or batteries<br />
incorrectly installed.<br />
AC power is connected, but no or<br />
weak batteries are installed.<br />
Batteries are weak or dead and no<br />
AC power source is connected.<br />
An External AC Power Supply cord<br />
is not connected to the monitor<br />
or accessory cable. The accessory<br />
cable is not connected to the<br />
monitor.<br />
The AC power cord is connected<br />
to a wall outlet that is controlled<br />
by a wall switch.<br />
A non user-serviceable fuse may<br />
have opened.<br />
The battery is missing.<br />
The battery is drained.<br />
The battery is defective.<br />
The battery is defective.<br />
The Li-Ion rechargeable battery<br />
pack must be replaced.<br />
Ensure the batteries are installed<br />
correctly.<br />
Connect to source of AC power.<br />
This is proper operation.<br />
Replace or recharge the batteries.<br />
Replace or recharge the batteries.<br />
Connect to source of AC power.<br />
Check all power supplies and<br />
cables for proper connections.<br />
Only connect the AC power cord to<br />
an outlet that is not controlled by a<br />
wall switch.<br />
Contact your authorized service<br />
representative.<br />
Replace the battery.<br />
Charge or replace the battery.<br />
Contact your authorized service<br />
representative if using LI-Ion<br />
rechargeable battery pack.<br />
Disconnect the AC power cord<br />
and then reconnect it. If the error<br />
message persists, the battery is<br />
defective. Contact your authorized<br />
service representative.<br />
Contact your authorized service<br />
representative.<br />