Chapter 11: Troubleshooting<br />

Power<br />

PROBLEM POSSIBLE CAUSE CORRECTIVE ACTION<br />

The oximeter doesn’t<br />

turn on.<br />

Batteries weak and/ or no AC<br />

power.<br />

Replace or recharge the batteries.<br />

Connect to source of AC power.<br />

The oximeter turns<br />

on, but display reads:<br />

L0<br />

Bat<br />

The oximeter turns<br />

off unexpectedly.<br />

The External Power<br />

indicator X is not<br />

lit (AC Power).<br />

The monitor operates<br />

on AC power, but not<br />

on battery power.<br />

The monitor displays<br />

27<br />

ERR when powered<br />

up.<br />

Battery run time is<br />

excessively short on a<br />

fully charged battery.<br />

Batteries not installed or batteries<br />

incorrectly installed.<br />

AC power is connected, but no or<br />

weak batteries are installed.<br />

Batteries are weak or dead and no<br />

AC power source is connected.<br />

An External AC Power Supply cord<br />

is not connected to the monitor<br />

or accessory cable. The accessory<br />

cable is not connected to the<br />

monitor.<br />

The AC power cord is connected<br />

to a wall outlet that is controlled<br />

by a wall switch.<br />

A non user-serviceable fuse may<br />

have opened.<br />

The battery is missing.<br />

The battery is drained.<br />

The battery is defective.<br />

The battery is defective.<br />

The Li-Ion rechargeable battery<br />

pack must be replaced.<br />

Ensure the batteries are installed<br />

correctly.<br />

Connect to source of AC power.<br />

This is proper operation.<br />

Replace or recharge the batteries.<br />

Replace or recharge the batteries.<br />

Connect to source of AC power.<br />

Check all power supplies and<br />

cables for proper connections.<br />

Only connect the AC power cord to<br />

an outlet that is not controlled by a<br />

wall switch.<br />

Contact your authorized service<br />

representative.<br />

Replace the battery.<br />

Charge or replace the battery.<br />

Contact your authorized service<br />

representative if using LI-Ion<br />

rechargeable battery pack.<br />

Disconnect the AC power cord<br />

and then reconnect it. If the error<br />

message persists, the battery is<br />

defective. Contact your authorized<br />

service representative.<br />

Contact your authorized service<br />

representative.<br />

11-2 Veterinary <strong>V1030</strong> <strong>Hand</strong>-<strong>Held</strong> <strong>Pulse</strong> <strong>Oximeter</strong> Operation Manual