Chapter 4: Operating Instructions<br />
The PI value is represented as a 9-segment bar graph. The more bars lit, the higher the PI value,<br />
and generally the better the sensor site. If only the first one or two segments are lit, the segment<br />
color changes to yellow, indicating a technical alert condition. This indicates that the oximeter is<br />
receiving a low signal quality from that sensor site, and further degradation of the signal quality<br />
could cause the oximeter to lose its ability to obtain readings. A different sensor site should be<br />
considered.<br />
The PI value approximately maps to a 9-segment bar graph as shown below. The two lowest bars<br />
(1 and 2) 2 are bi-color (yellow and green), and bars 3 through 9 are green.<br />
10.24 - 20.47<br />
5.12 - 10.23<br />
2.56 - 5.11<br />
1.28 - 2.55<br />
0.64 - 1.27<br />
0.32 - 0.63<br />
0.16 - 0.31<br />
0.08 - 0.15<br />
0.00 - 0.07<br />
NOTE! The PI value is a relative value that varies from patient to patient.<br />
4-12 Veterinary <strong>V1030</strong> <strong>Hand</strong>-<strong>Held</strong> <strong>Pulse</strong> <strong>Oximeter</strong> Operation Manual