Location SelectionSelect a location for the instrument that is:! Firm and level,! Free from vibration that will affect the balance measurements,! Close to an isolated power outlet that will provide dedicated electrical power, and! Away from the direct flow of a fan or heating/air-conditioning outlet that mightproduce lifting force on the balance.Install the pan support.Connect the nitrogen purge lines if used (page 15).Connect a printer if used (page 17).Connect a computer if used (page 20).Ensure the AC power receptacle is set for the correct voltage (120 or 240 VAC). (page13)The <strong>MAX</strong>-<strong>1000</strong> will draw up to seven (7) amps. Ensure the instrument is plugged intoits own DEDICATED and GROUNDED electrical power outlet. Do not put it on acircuit with motors, blenders, heaters, coolers, grinders, or other high current electricaldevices. However, a low current (less than one amp) device would be safe.Turn the instrument on by pressing the switch on the back of the analyzer above thepower receptacle.Adjust the contrast as necessary to view the screen display.The <strong>MAX</strong>-<strong>1000</strong> is now ready for use.3