12.1.9 If this is the first test of the day, discard the results of this test for stability reasons.12.1.10 Record the final result. The value should fall between 37.48 and 37.52.12.1.11 If the results do not meet this criteria a weighing error has occurred.12.1.12 The most common causes of such errors are bent or dirty sample pans, ordebris falling into the balance shaft interfering with the pan support’s ability tofreely move up and down.12.1.13 After the instrument is cleaned:! Use a new sample pan and repeat the weight test five more times.! Calculate the results.! If the results are still outside the specified range, call AZI Customer Service forassistance.12.2 PRECISION WEIGHT TEST12.2.1 Run five weight tests and record the results.12.2.2 The recorded results should fall within the following.NUMBER OFSAMPLES5MEAN 37.48 TO 37.52STANDARDDEVIATIONCOEFFICIENT ofVARIATIONLESS THAN 0.01LESS THAN 0.03%12.3 SODIUM TARTRATE TESTThe sodium tartrate test verifies the heater and force balance operation. Sodiumtartrate dehydrate is a chemical with a known moisture concentration which givesup moisture at temperatures greater than 120 o C.Because of its consistent moisture concentration, it is used as a standard inmoisture measurement using oven or titration methods. To obtain consistently56