3.6.4 After “START” has been pressed, a weight error message may appear on thisscreen. For example, if no pan or pan support is on the balance, UNDERLOADwill be displayed. If the weight on the sample pan exceeds 40 grams,OVERLOAD will be displayed. If an empty sample pan is in place, and either ofthese displays appear, turn the unit off , remove the sample pan, pan support, thecircular balance shield and clean any debris found beneath the pan support.Reassemble the parts and be sure that no mechanical interference is present. Itmay be necessary to calibrate the balance (see page 37). If these measures fail torestore normal function, call AZI Customer Service at (800) 235-3360 or (602)470-1414 for assistance.9