If smaller standard deviations are needed:<strong>MAX</strong>-<strong>1000</strong> - Mid-range predictIf improved standard deviations are needed:<strong>MAX</strong>-<strong>1000</strong> - Tight predictIf standard deviations are still not acceptable, try RATE ending criteria.SAMPLE SIZE: Adjust as necessary for application. Smaller sample sizes of ahomogenous material will decrease test times.TARE OPTIONS:Pan tare:StandardSample tare:When stable. Use sample tare “3 seconds” with warm orhot applications.Ultra-low pan tare: With samples having moisture levels below 1%PARAMETER DOCUMENTATION:To ensure optimum and efficient correlation, reproducibility, and for record keepingpurposes, record the following.- Standard testing - 3 Karl Fischer, 3 oven tests or other accepted reference methodprocedure.- 5 repeat tests on the <strong>MAX</strong>-<strong>1000</strong>.- Record comments on special techniques required to duplicate accurate andreproducible results.If you need assistance with parameter development or have questions concerning theoperation of the analyzer, please contact either your Sales Representative or CustomerService Representative at (800) 235-3600 or (602) 470-1414.73