Computrac MAX-1000 User's Manual - Arizona Instrument
Computrac MAX-1000 User's Manual - Arizona Instrument
Computrac MAX-1000 User's Manual - Arizona Instrument
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
TEMPERATURE CIRCUITProbable cause:Action:A/D TIMEOUTProbable cause:Action:INTERRUPT STARTUPProbable cause:Action:Message is generated if temperature circuit is not detected.A-to-D failure. RTD failure. Bridge circuit failure. MPUfailure.Troubleshoot temperature circuit. Check RTD connection toboard.Message generated if A/D does not clear in 100mS.Bad A/D. Bad A/D timing circuit.Troubleshoot A/D circuit.Something caused the system to crash and reset.Usually RAM corruption of some kind.Troubleshoot system.65