The low input reading is referenced to 60mV <strong>and</strong> is calculated using this equation: (see section 3.3.6)(Sensor Output @ Low PV)LOW RDG = (1000) * ------------------------( 60mV )Press YES to select the menu or NO to advance to the next menu.Press UP <strong>and</strong> DOWN to set the low reading.Press ENTER to store the value or BACK to return to the SETUP LOOP INPUT main menu prompt.Linear Input Scaling Example:PROBLEM: We want to connect a pressure sensor to enable us to read pressure directly in PSI. The sensor has anoutput of 4-20mA over a range of 100-400PSI. The 4-20mA is converted to 12-60mV (see section 3.3.6 <strong>and</strong> 3.3.7)d.How do we set the input scaling?SOLUTION: 1. Set the HIGH PV to 400 <strong>and</strong> enter the units as P, S,<strong>and</strong> I.2. Calculate the HIGH RDG <strong>and</strong> enter it:(60mV)