<strong>Plum</strong> A+ <strong>Infusion</strong> Systems 6- 1322) Select [ENTER] to keep change(s) <strong>and</strong> returnto the Main Delivery screen, or [CANCEL/BACK] to restore previous values <strong>and</strong> return tothe Options screen.6.7 Variable Rate CapVariable Rate Cap is a biomed programmable item to set themaximum rate values to be accepted by the device. The cliniciancannot change it once biomed has programmed it.6.8 Examples of Automatic CalculationAT STARTUPInitial programming allows the clinician to enter two of the threeprogramming parameters (Rate, VTBI, or Duration) <strong>and</strong> the thirdis automatically calculated. (Refer to table below)1st Action 2nd Action [AUTOCALC]enter RATE enter VTBI [DURATION]enter VTBI enter DURATION [RATE]enter RATE enter DURATION [VTBI]WHILE RUNNING (TITRATION)Changing two parameters after startup of the infusion will allowfor the re-calculation of the third parameter.430-95215-003 (Rev. 10/05)