9- 4 9) SpecificationsTIME TO DETECT DOWNSTREAM OCCLUSIONSRATEOCCLUSIONALARMPRESSURESETTINGTYPICAL TIMETO ALARM(MACRO)25 mL/hr6 psig15 psig38 seconds98 seconds1 mL/hr 6 psig17 minutes15 psig45 minutes* Baseline backpressure is 0 psig *MAXIMUM TIMETO ALARM(MACRO)47 seconds106 seconds22 minutes49 minutesBOLUS VOLUME RELEASED AFTERDOWNSTREAM OCCLUSIONS ARE CORRECTEDRATEOCCLUSIONALARMPRESSURESETTINGTYPICALBOLUSVOLUME(MACRO)25 mL/hr6 psig15 psig0.23 mL0.55 mL1 mL/hr 6 psig0.24 mL15 psig 0.59 mL* Baseline backpressure is 0 psig *MAXIMUM BOLUSVOLUME(MACRO)0.28 mL0.63 mL0.30 mL0.70 mLDELIVERY ACCURACYThe <strong>Plum</strong> A+ Infuser was designed <strong>and</strong> manufactured to maintain avolumetric delivery rate error of the total fluid delivered of less thanor equal to ± 5% over the course of 48 hours at a programmed rateof 1 to 999 mL/hr during normal operating conditions. For use of thedevice at rates below 1 mL/hr, the delivery rate error is less than orequal to ± 10%.Backpressure Effect- Positive backpressure on the distal line mayaffect delivery accuracy by no more than 0.5% per psig (2% per430-95215-003 (Rev. 10/05)