<strong>Plum</strong> A+ <strong>Infusion</strong> Systems 7- 17) Alarms <strong>and</strong> Troubleshooting7.1 Warning MessagesMESSAGEStop delivery then turnoffWarning: ReplaceBatteryWarning: ChargerServiceWarning: Low BatteryPOSSIBLE CAUSEAttempting to turn theinfuser off while adelivery is in progress.The battery currentcharge is greater than200 mA/hr.The battery voltage isless than the depletedthreshold <strong>and</strong> chargelevel is higher than thelow charge threshold.The charging circuitryis not behaving asexpected (chargevoltage is notchanging) <strong>and</strong> canreduce battery life.Infuser not plugged into AC power.CORRECTIVEACTIONStop all lines, then turninfuser off.Battery needs to bereplaced.Silence key is pressed.Plug into AC power,only about 30 minutesof battery life remains.NOTE: Other error <strong>and</strong> warning messages will occurwhen appropriate for the current selection context. Forinstance, when a change to a required parameter wouldresult in clearing some delivery values, the warningmessage will appear when that field is selected. When aninput error has just been made, the three quick beeps ofan Invalid Key Warning will sound <strong>and</strong> an appropriate430-95215-003 (Rev. 10/05)