Abbott Hospira Plum A Plus and Plum A Plus 3 version 11.6 Infusion ...
Abbott Hospira Plum A Plus and Plum A Plus 3 version 11.6 Infusion ...
Abbott Hospira Plum A Plus and Plum A Plus 3 version 11.6 Infusion ...
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
4- 20 4) Basic Operation4.9 BackprimingBACKPRIMING- Removal of airor fluid from the proximal line<strong>and</strong> cassette airtrap.1) To initiate a backprime,select [BACK PRIME].ASTOPPED STOPPEDRate5.50mL/hrVol Inf8.9 0mLBNOTE: Backpriming is availableonly when all delivery is stopped.Secondary container is necessaryto receive the air.BackPrimeABOptions/Vol Inf2) Select <strong>and</strong> hold [BACK PRIME] until fluidpumped from line A toline B has expelled the airfrom cassette air trap intoSecondary container.NOTE: Fluid flow is at maximumrate possible.ABACKPRIMINGRate5.50mL/hrVol Inf8.9 0mLRelease Backprime to StopBackPrimeABBOptions/Vol Inf3) Upon release of [BACK PRIME], the cassettetest is performed to confirm readiness forpumping. Pressing will resumeprogrammed delivery.430-95215-003 (Rev. 10/05)