4- 4 4) Basic OperationFailure during the Self-Tests is reported as a MalfunctionCondition.DATA RETENTIONDelivery program settings <strong>and</strong> programming option selections areretained in non-volatile memory. If the <strong>Plum</strong> A+ has been turnedoff for longer than four hours, all delivery settings are cleared <strong>and</strong>programming option selections are restored to their defaultselections for next use.4.2 Power On1) Press to turn on the<strong>Plum</strong> A+ <strong>and</strong> you will seethe System Self-Testsscreen.CAUTION: THE INFUSER MUST BEINITIALLY STARTED ON AC POWER,OTHERWISE THE DATAPORT MAY NOTOPERATE PROPERLY.After self-testingsuccessfully completes,the display will proceed tothe next screen.HOSPIRA <strong>Plum</strong> A+Release XX.XXCopyright <strong>Hospira</strong>2005System Self TestIn Progress**SCREENS SHOWN IN THIS MANUALARE EXAMPLES ONLY AND DON’TNECESSARILY REFLECT ACTUALCLINICAL PRACTICE**Programming the infuser may begin during thecassette check phase of the start-up testing.NOTE: If battery is missing or defective, a “Battery notinstalled” message with a “Continue without battery” promptwill appear (AC power only).430-95215-003 (Rev. 10/05)