7- 4 7) Alarms <strong>and</strong> TroubleshootingCODE MESSAGE POSSIBLE CAUSEN250N251N252N253N254N255N256Door openwhile pumpingCassette TestFailureDepletedBatteryHard lockoutViolationHard lockoutEnabledSoft lockoutViolationSoft lockoutEnabledPump door openedwhile operating.Faulty cassette,proximal or distalocclusion or air incassette detected atstart up.Too much time onbattery power.The use of the [STOP]key during delivery, oropening the door,while lockout isenabled.Pressing of any keyexcept [STOP] duringdelivery while anyalarm is active <strong>and</strong>lockout is enabled.The use of the [STOP]key during delivery, oropening the door,while lockout isenabled.Pressing of any keyexcept [STOP] duringdelivery while anyalarm is active <strong>and</strong>lockout is enabled.CORRECTIVEACTIONClose door withcassette inserted.Check clamps,manually reprime set,close door.Backprime into line Btubing or syringe ifappropriate. If alarmrecurs on retest,replace set. If alarmrepeats, replaceinfuser.Plug into AC (mains)power.Unlock lockoutswitch.Unlock lockoutswitch.Unlock lockout bypressing the decimalkey, followed by 9, 6,& 3 on the keypad.Unlock lockout bypressing the decimalkey, followed by 9, 6,& 3 on the keypad.430-95215-003 (Rev. 10/05)