4- 10 4) Basic OperationThe following 3 screens are examples ofchanged values.Example of changed rate.APROGRAM4) From the Main DeliveryScreen, select [A].RateVTBI1550mL/hrmL5) Use keypad to change theRATE value.Duration03:20Enter Value using keypadProgramTherapyOptionshr : minCancel/BackNOTE: The display time is onlyaccurate to ± one minute. Asshown in this example, the delivery shows 3 hours <strong>and</strong> 20minutes (not 3 minutes <strong>and</strong> 20 seconds).NOTE: [START] accepts the change. Delivery is changed tothe new RATE <strong>and</strong> DURATION <strong>and</strong> display returns to MainDelivery Screen.Example of Changed VTBIvalue.6) Use key to highlightthe VTBI field.A PROGRAMRateVTBIDuration1510006:40mL/hrmLhr : min7) Use the keypad to changethe VTBI value.Enter Value using keypadProgramTherapyOptionsCancel/BackNOTE: When the VTBI is changed,the Duration is recomputed with respect to the current Rate.To underst<strong>and</strong> the Automatic Calculation feature, see Section6.8 Examples of Automatic Calculation.NOTE: [START] accepts the change. Delivery is changed tothe new VTBI <strong>and</strong> DURATION <strong>and</strong> display returns to MainDelivery Screen.430-95215-003 (Rev. 10/05)