7- 8 7) Alarms <strong>and</strong> TroubleshootingCODEN182,N188,N189MESSAGEProx Occl B/ AirN231 Prox Air B,BackprimeMALFUNCTIONSPOSSIBLECAUSEProximal lineobstructed, line Bcontainerdisconnected orobstructed.When proximal airdetected exceeds thelimit for the line.CORRECTIVEACTIONExamine line B forkinks, closed clamps,or air in cassette.Occlusion caused byempty container,backprime into line Btubing or syringe.Check for clamps, orempty container.Correct as necessary.Backprime into line Btubing or syringe.CODEMESSAGEPOSSIBLECAUSECORRECTIVEACTIONE300 orhigherMalfunctionA failure has occurredin the infuser’sinternal systems.Record Malfunctionerror code. Pumpmust be turned OFFto clear malfunction.If alarm malfunctionrepeats, replaceinfuser.430-95215-003 (Rev. 10/05)