<strong>Plum</strong> A+ <strong>Infusion</strong> Systems 7- 7LINE B ALARMSCODEMESSAGEPOSSIBLECAUSECORRECTIVEACTIONN104CallbackLine BAlarm programmedby user. Change indelivery for piggybackdelivery, LoadingDose to MaintenanceStep, or for any stepin “Multistep” therapyprovided the step isnot the last step.Press [SILENCE].N160Line B VTBIcompleteVTBI complete on lineB.Program new VTBI.N183Prox Occl BAt StartupProximal lineobstructed, line Bcontainer (or syringe)disconnected orobstructed duringnon-delivery (startup).Examine line B forkinks or closed clamps.If there are closedclamps, open theclamps. Then, open<strong>and</strong> close door.Continueprogramming. Replaceset if problem persists.If clamps are notopened prior tocompletion of cassettetest, peak proximalocclusion may recur.For syringe, select [Backprime].430-95215-003 (Rev. 10/05)