6- 14 6) Additional FeaturesNOTE: VTBI will not be re-calculated if the Rate <strong>and</strong> Durationare changed. VTBI must be cleared <strong>and</strong> then the new VTBIre-entered. The new VTBI will automatically calculate a newRate.1st Action 2nd Action [AUTOCALC]change RATE keep VTBI [DURATION]change RATE change VTBI [DURATION]change VTBI keep RATE [DURATION]change DURATION keep VTBI [RATE]change VTBI change DURATION [RATE]change RATE change DURATION Recalculates[RATE] based onprevious VTBI <strong>and</strong>new DURATIONAT KVOAudible KVO alarm will sound, clinician will need to re-set theappropriate settings.1st Action 2nd Action [AUTOCALC]enter VTBI keep RATE [DURATION]enter DURATION keep RATE [VTBI]change RATE enter VTBI [DURATION]enter DURATION enter VTBI [DURATION]enter VTBI enter DURATION [RATE]change RATE change DURATION [VTBI]430-95215-003 (Rev. 10/05)