<strong>Plum</strong> A+ <strong>Infusion</strong> Systems 3- 9Insert piercing pin into container outlet witha twisting motion.Fill Drip chamber to about 1/2 full or to thescore mark. Do not completely fill.NOTE: To avoid getting Distal Air Alarms,confirm the fluid bag is positioned higherthan the infuser. This should be set upprior to priming the cassette.PRIMING THE ADMINISTRATION SETInvert the cassette.Turn the flow regulator counter-clockwiseuntil a drop of fluid is seen in the pumpingchamber.Turn the cassette upright, then prime theremainder of the administration set.430-95215-003 (Rev. 10/05)