<strong>Plum</strong> A+ <strong>Infusion</strong> Systems 4- 15Verify change to“Concurrent” mode.2) Use to highlight Ratefield.ModeRateVTBIDurationProgramOptionsPROGRAMConcurrent0000:00Change using Change ModeChangeModeTherapyBmL/hrmLhr : minCancel/Back3) To enter delivery rate, usenumeric keypad.NOTE: When Rate for Line A plusRate for Line B is greater than validrate ranges, a “Concurrencyviolation” message will explain whysecond line did not start. SeeSection 9.3 for valid delivery rateranges.ModeRateVTBIDurationPROGRAMConcurrent10000:00Enter value using keypadProgramTherapyOptionsBmL/hrmLhr : minCancel/BackNOTE: “Concurrency violation” may not be displayed if unit isdisplaying “VTBI Complete”. The device will not start <strong>and</strong> aninvalid keypress sound will occur when START is pressed.4) Use key to move down to the VTBI field.5) To enter VTBI amount,use keypad.NOTE: While the VTBI is beingentered, the duration isautomatically computed. While thecomputed value is greater than99:59, dashes (- -) appear in placeof hours. To underst<strong>and</strong> theModeRateVTBIDurationPROGRAMConcurrent1010010:00Enter value using keypadProgramTherapyOptionsBmL/hrmLhr : minCancel/Back430-95215-003 (Rev. 10/05)