Abbott Hospira Plum A Plus and Plum A Plus 3 version 11.6 Infusion ...
Abbott Hospira Plum A Plus and Plum A Plus 3 version 11.6 Infusion ...
Abbott Hospira Plum A Plus and Plum A Plus 3 version 11.6 Infusion ...
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
3- 10 3) Equipment DescriptionPush in the flow regulator to close. Confirmthere is no flow.LOADING THE CASSETTETo load the primed cassette into the plum A+, proceed as follows.Open the cassette door by lifting up theh<strong>and</strong>le.Insert cassette into door guides. Close thecassette door. Confirm there is no flow <strong>and</strong>no kinks appear in tubing.NOTE: If flow is observed, close the tubingclamps <strong>and</strong> replace the administration set.NOTE: The administration set should bechanged per CDC guidelines or healthcareprovider policy. Discard after use.PREPARING THE SECONDARY LINECAUTION: CONSULT THE DRUG LABELING TO CONFIRMDRUG COMPATIBILITY, CONCENTRATION, DELIVERY RATES,AND VOLUMES ARE ALL SUITABLE FOR CONCURRENTDELIVERY OR PIGGYBACK DELIVERY (SECONDARY FOLLOWEDAUTOMATICALLY BY PRIMARY) MODES.The <strong>Plum</strong> A+ features concurrent or piggyback delivery modeswhen therapy requires administering more than one fluid.430-95215-003 (Rev. 10/05)