9- 2 9) SpecificationsRecharge:Electrical Leakage:NURSE-CALLSystem:Circuitry Ratings:The battery charges whenever the infuser(s) isconnected to AC power. The recharge time isapproximately six hours with the deviceoperating at 125 mL/hr on one line.Meets ANSI/AAMI ES1-1993 st<strong>and</strong>ard: SafeCurrent Limits for Electro-medical Apparatus.NURSE-CALL alarm is factory set for Normally-Open (NO)Contact the Technical Services Center to makean internal adjustment to change the devicefrom Normally-Open (NO) to Normally Closed(NC) system.Voltage-30 VDC MaxCurrent- 0.25 Amps MaxContact Rating- 3 Watts MaxVTBI RANGEVTBI Range:ENVIRONMENTOperatingTemperature:Storage Temperature:AtmosphericPressure:Relative Humidity:0.1 to 99.9 mL (in 0.1 mL increments)100 to 9999 mL (in 1 mL increments)5° to 40° C-20° to 60° C0 - 10,000 feet (0 - 3,000m) or equivalentpressure10 - 90% (40° C Max)DELIVERY RATE RANGELines A <strong>and</strong> B: 0.1 - 99.9 mL/hr (in 0.1 mL increments)100 - 999 mL/hr (in 1 mL increments)430-95215-003 (Rev. 10/05)