12.07.2015 Views

Abbott Hospira Plum A Plus and Plum A Plus 3 version 11.6 Infusion ...

Abbott Hospira Plum A Plus and Plum A Plus 3 version 11.6 Infusion ...

Abbott Hospira Plum A Plus and Plum A Plus 3 version 11.6 Infusion ...

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>Plum</strong> A+ <strong>Infusion</strong> Systems 7- 5LINE A ALARMSCODE MESSAGE POSSIBLE CAUSEN105N161N185Callback LineALine A VTBIcompleteProx Occl A AtStartupAlarm programmed byuser. Change indelivery for LoadingDose to MaintenanceStep or for any step in“Multistep” therapyprovided the step isnot the last step.VTBI complete on lineA.Proximal lineobstructed, line Acontainer (or syringe)disconnected orobstructed during nondelivery(startup).CORRECTIVEACTIONPress [SILENCE].Program new VTBI.Examine line A forkinks or closedclamps. If there areclosed clamps, openthe clamps. Then,open <strong>and</strong> close door.Continueprogramming.Replace set ifproblem persists.If clamps are notopened prior tocompletion ofcassette test, peakproximal occlusionmay recur.For syringe, select [Backprime].430-95215-003 (Rev. 10/05)