4- 12 4) Basic OperationThe Delivery Mode field ishighlighted.NOTE: If an entry was previouslyprogrammed, RATE is highlighted.If no previous entries, MODE ishighlighted.2) If field reads“Concurrent”, select [CHANGE MODE] toswitch to “Piggyback”.ModeRateVTBIDurationProgramOptionsPROGRAMPiggyback0000:00Change using Change ModeChangeModeTherapyBmL/hrmLhr : minCancel/Back3) Select to highlight theRATE field.4) To enter delivery rate, usenumeric keypad.5) Use key to movedown to the VTBI field.ModeRateVTBIDurationPROGRAMPiggyback200000:00Enter value using keypadProgramTherapyOptionsBmL/hrmLhr : minCancel/Back6) To enter VTBI amount,use keypad.NOTE: While the VTBI is beingentered, the duration isautomatically computed. If thecomputed value is greater than99:59, dashes (- -) appear in placeof hours. To underst<strong>and</strong> theAutomatic Calculation feature, seeSection 6.8 Examples of AutomaticCalculation.ModeRateVTBIDurationPROGRAMPiggyback20010000:30Enter value using keypadProgramTherapyOptionsBmL/hrmLhr : minCancel/Back430-95215-003 (Rev. 10/05)