3- 12 3) Equipment DescriptionSet up per operating instructions. (SeeSections 4 - 6 of this manual.)Syringe: Secure the container to thecassette door using the optional containersupport arm (List 12095-03).3.5 Discontinuing Electronic Flow Control& Setting Gravity FlowCAUTION: BEFORE OPENING DOOR, CLOSE CLAMP ONPRIMARY OR SECONDARY SET OR REMOVE SECONDARYCONTAINER FROM SECONDARY PORT TO PREVENT MIXING OFPRIMARY AND SECONDARY FLUIDS.If electronic flow needs to be stopped, the precision flow regulatoron the cassette can be used to manually set <strong>and</strong> maintain fluidflow in a gravity mode (see the following graphics).1PRESS STOPSTOP2 OPEN DOOR & REMOVE CASSETTENOTE: FLOW REGULATOR CLOSESAUTOMATICALLY WHEN DOOR IS OPENEDTO PREVENT ACCIDENTAL FREE FLOW3 SET GRAVITY RATE BYTURNING FLOW REGULATORCOUNTER-CLOCKWISENOTE: CASSETTE SHOULD BEIN UPRIGHT POSITIONNOTE: IF DOOR OPEN ALARM OCCURS, TURN PUMP OFF430-95215-003 (Rev. 10/05)