<strong>Plum</strong> A+ <strong>Infusion</strong> Systems 9- 3Concurrent Delivery:<strong>Plum</strong>Set:KVO:AIR-IN-LINE ALARM<strong>Plum</strong>Set (Distal):<strong>Plum</strong>Set (Proximal):0.5 mL/hr minimum for each line500 mL/hr cumulative (A+B) maximum1.0 mL/hr or the last primary delivery rate,whichever is lessBolus 0.1 mL (100 microliters) or larger fordelivery rates less than 500 mL/hr.Bolus 0.5 mL (500 microliters) or larger fordelivery rates equal to or greater than 500mL/hr.Cumulative 0.25 mL out of 2.6 mL fordelivery rates less than 500 mL/hr.Cumulative 0.50 mL out of 5.3 mL fordelivery rates equal to or greater than 500mL/hr.Bolus at 0.5 mL, Total 1.0 mLOCCLUSION ALARM AND LIMITSDistal Occlusion:Proximal Occlusion:Distal Pressure Limit(without alarm):Maximum <strong>Infusion</strong>Pressure:The DISTAL OCCLUSION alarm soundsafter the distal set tubing or set outlet fittingbecomes occluded.The PROXIMAL OCCLUSION alarmsounds if the tubing proximal to thecassette becomes occluded.1 to 15 psig. The maximum pressure isuser-selectable. Factory default setting is 6psig.20 psig430-95215-003 (Rev. 10/05)