<strong>Plum</strong> A+ <strong>Infusion</strong> Systems 5- 25NOTE: When Step-4 isprogrammed, using the [SELECT]keys to advance will highlightanother item on this screen.13) When Step-4 isprogrammed, select [MORE STEPS] toprogram up to 10 Steps.ATPN1234Program MultistepEnter value using keypadProgramOptionsRatemL/hr501007550VTBImL508257550MoreStepsDurationhr : min01 : 0008 : 1501 : 0001 : 00Cancel/BackNOTE: Using the [SELECT] keyswill not advance you to theadditional steps.14) To set up Steps 5 through10, enter desired valuesas performed for Steps 1through 4.To see the previous steps,select [PREV STEPS].NOTE: Confirmation screen willappear ONLY if using weightbasedprogram.ATPN5678910Program MultistepEnter value using keypadProgramOptionsRatemL/hr000000VTBImL000000PrevStepsDurationhr : min00 : 0000 : 0000 : 0000 : 0000 : 0000 : 00Cancel/BackCAUTION:INFUSION.VERIFY ALL VALUES BEFORE STARTING15) Press to begin program.430-95215-003 (Rev. 10/05)