4- 14 4) Basic Operation4.6 Concurrent DeliveryCONCURRENT DELIVERY-Simultaneous delivery of twofluids at independent flow rates.CAUTION: CONSULT DRUGLABELING TO CONFIRM DRUGCOMPATIBILITY, CONCENTRATION,DELIVERY RATES, AND VOLUME AREALL SUITABLE FOR CONCURRENT ANDPIGGYBACK DELIVERY MODES. SEESECTION 8 FOR INFORMATION ONCONCURRENT FLOW, SEE SECTION 9FOR DELIVERY RATE RANGES.APUMPINGASTOPPEDRate5.50mL/hrVol Inf0.1 0mLBBOptions/Vol InfNOTE: In concurrent mode with certain alarm conditions, oneline may stop while the other line may continue to deliver.1) With pumping started on line A, to program aConcurrent delivery rate, select [B].The Delivery Mode field ishighlighted.NOTE: If an entry was previouslyprogrammed, RATE is highlighted.If no previous entries, MODE ishighlighted.If field reads “Piggyback”,select [CHANGE MODE] toswitch to “Concurrent”.ModeRateVTBIDurationProgramOptionsPROGRAMPiggyback0000:00Change using Change ModeChangeModeTherapyBmL/hrmLhr : minCancel/Back430-95215-003 (Rev. 10/05)