09.10.2015 Views

OS-C501

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Offshore Standard DNV-<strong>OS</strong>-<strong>C501</strong>, November 2013<br />

Sec.10 Component testing – Page 162<br />

3.1.9 If analysis and test results agree with each other based on the criteria given in [3.2] and [3.3] the analysis<br />

method is suitable for the application. (See also Sec.7 [2.2]).<br />

3.2 Short term tests<br />

3.2.1 The requirements here apply to one test aspect as determined in [3.1].<br />

3.2.2 The sequence of the failure modes in the test shall be the same as predicted in the design. If the sequence<br />

is different or if other failure modes are observed, the design shall be carefully re-evaluated.<br />

3.2.3 The measured strength of each critical failure mode shall never be less than the predicted characteristic<br />

strength. Critical failure modes are failure modes that are linked to a limit state. The characteristic strength of<br />

the component shall be updated according to [3.3].<br />

3.2.4 In addition to the requirements above ([3.2.1] and [3.2.2]), one of the following requirements shall be<br />

met:<br />

— the test results fulfil the requirements for “confirmation testing for static data” given in Sec.4 [8.7].<br />

Application of this criterion requires that at least three tests are carried out<br />

— the characteristic strength of the structure is updated by the test results as described in [3.4]. One test is<br />

sufficient to use this method, but more tests are recommended.<br />

3.3 Long term testing<br />

3.3.1 Whether cyclic load testing or long term static load testing or both is required depends on the evaluation<br />

of the test programme done in [3.1]. The approach for both testing types is similar and will be treated here in<br />

one part.<br />

3.3.2 Fatigue testing for high safety class: at least two survival tests shall be carried out. The specimen should<br />

not fail during the survival test and it should not show unexpected damage. The requirements to the testing are:<br />

— tests should be carried out up to five times the maximum number of design cycles with realistic amplitudes<br />

and mean loads that the component will experience. If constant amplitude testing is carried out tests should<br />

be carried out up to 50 times the maximum number of design cycles to compensate for uncertainty in<br />

sequence effects.<br />

— if the anticipated lifetime exceeds 10 5 cycles testing up to 10 5 cycles may be sufficient. The load levels<br />

should be chosen such that testing of the two specimens is completed after at least 10 4 and 10 5 cycles<br />

respectively. The logarithms of the two test results shall fall within µ-σ of the logarithm of the anticipated<br />

number of cycles to failure, where µ is the mean of the logarithm of the predicted number of cycles to<br />

failure and σ is one standard deviation of the logarithm of the predicted number of cycles to failure, both<br />

interpreted from a log(stress)-log(lifetime) diagram for the anticipated number of cycles to failure. If more<br />

tests are made the requirements are given in DNV-<strong>OS</strong>-<strong>C501</strong> Sec.4 [8.8.6].<br />

3.3.3 Fatigue testing for normal safety class: at least one survival test shall be carried out. The specimen should<br />

not fail during the survival test and it should not show unexpected damage. The requirements to the testing are:<br />

— tests should be carried out up to three times the maximum number of design cycles with realistic amplitudes<br />

and mean loads that the component will experience. If constant amplitude testing is carried out tests should<br />

be carried out up to 30 times the maximum number of design cycles to compensate for uncertainty in<br />

sequence effects.<br />

— if the anticipated lifetime exceeds 10 5 cycles testing up to 10 5 cycles may be sufficient. The load levels<br />

should be chosen such that testing of the two specimens is completed after at least 10 4 and 10 5 cycles<br />

respectively. The logarithms of the two test results shall fall within µ-2σ of the logarithm of the anticipated<br />

number of cycles to failure, where µ is the mean of the logarithm of the predicted number of cycles to<br />

failure and σ is one standard deviation of the logarithm of the predicted number of cycles to failure, both<br />

interpreted from a log(stress)-log(lifetime) diagram for the anticipated number of cycles to failure. If more<br />

tests are made the requirements are given in DNV-<strong>OS</strong>-<strong>C501</strong> Sec.4 [8.8.6].<br />

3.3.4 Stress rupture testing for high safety class: at least two survival tests shall be carried out. The specimen<br />

should not fail during the survival test and it should not show unexpected damage. The requirements to the test<br />

results are:<br />

— tests should be carried out up to five times the maximum design life with realistic mean loads that the<br />

component will experience. If constant load testing is carried out tests should be carried out up to 50 times<br />

the design life to compensate for uncertainty in sequence effects.<br />

— if the anticipated lifetime exceeds 1000 hours testing up to 1000 hours may be sufficient. The load levels<br />

should be chosen such that testing is completed after 10 3 hours. The logarithms of the two test results shall<br />

fall within µ-σ of the logarithm of the anticipated lifetime, where µ is the mean of the logarithm of the<br />

predicted lifetime and σ is one standard deviation of the logarithm of the predicted lifetime, both interpreted<br />

DET NORSKE VERITAS AS

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!