04.11.2019 Views

SHRP-A-379 traducir

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

of thermal expansion of the asphalt binder and the inserts is necessary to reduce thermal<br />

shrinkage stresses at the interface that otherwise cause bond failures.<br />

NOTE1.--PMMA (Plexiglas®), meeting MIL STD-5545, has been found acceptable for making the inserts.<br />

It is readily available, opaque to the laser beam, has a satisfactory coefficient of thermal expansion, and is easy to<br />

machine.<br />

6.1.3 Environmental Conditioning and Testing Chamber--The environmental<br />

conditioning and testing chamber shall be capable of controlling temperature and humidity<br />

and performing isothermal conditioning of test specimens prior to testing. It should be on the<br />

order of 300 mm wide by 200 mm deep by 450 mm high and shall completely enclose the<br />

specimen and gripping system. It shall be capable of controlling test temperatures between<br />

-40°C and ambient with variations within the chamber _< 0.2°C. It shall be equipped with<br />

a cooling system that has the capacity to reduce the chamber temperature from ambient to<br />

-40°C within 40 minutes and to change the chamber temperature from -30 to -40°C or<br />

from -40 to -30°C within 10 minutes. Mechanical cooling or liquid nitrogen may be used<br />

to cool the chamber. It shall have a dehumidifying system with a capacity such that the<br />

formation of frost on the interior of the chamber, the test specimen, or any of the test<br />

fixtures is eliminated.<br />

6.1.3.1 The chamber shall be capable of storing a minimum of 16 test specimens on<br />

a rack which is thermally isolated from the walls and floors of the chamber such that heat<br />

conducted from the walls and floors of the chamber does not affect the temperature of the<br />

stored specimens.<br />

NOTE2.--A demolded test specimen is placed on Plexiglas ®, Teflon ®or other plastic plate (with<br />

approximate dimensions of 6 mm thick by 20 mm wide by 100 mm long) and transferred to the environmentally<br />

controlled test chamber for thermal conditioning. The use of a transfer plate will minimize deformation during<br />

handling. Each plate, when the above dimensions are used, accommodates up to four demolded test specimens.<br />

6.1.3.2 The chamber shall be fitted with a front-opening door for maintenance and<br />

standardization purposes and an access port that allows for insertion of the operator's hand<br />

and forearm to position test specimens on the storage shelf for conditioning and to position<br />

test specimens on the grips for testing. The access port shall be designed so that changes in<br />

chamber temperature are < 0.2°C during an operation in which the operator's hand or<br />

forearm is inserted into or removed from the chamber. Visual access to the interior of the<br />

test chamber shall be provided to permit proper mounting of test specimens and test<br />

monitoring. The test specimen elongation is measured with an optical laser. Use of the laser<br />

requires optical glass windows on two sides of the temperature chamber so that a beam of<br />

laser light can be passed through the chamber without distorting the beam.<br />

6.1.4 Load Measuring and Recording Devices--Load shall be measured with a load<br />

cell having a minimum capacity of 500 N and a sensitivity of 0.1 N. The load cell shall be<br />

calibrated at least annually in accordance with ASTM E4. The load and elongation shall be<br />

monitored with the data acquisition system such that they can be resolved to 1% of the<br />

failure load and elongation, respectively. Once the test has started, the data acquisition<br />

system shall be able to detect the point in time when the load starts to change as a result of<br />

elongation in the sample. This shall be accomplished by monitoring the load cell signal with<br />

47

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!