– 114 –<br />

NOTE<br />

In general, avoid any direct contact between mucosa and contra-angle as this can lead to<br />

false readings.<br />

9. Make the contra-angle slightly wet with a disinfecting wipe, for example and slip the<br />

contra-angle sleeve carefully over the contra-angle.<br />

Fig. 3 Contra-angle sleeve<br />

10. Disinfect the keypad and micromotor before first use and before each use on a new<br />

patient (for details see chapter 6 “Cleaning, disinfection”).<br />

2.3 Battery<br />

<strong>VDW</strong>.GOLD is operated by a rechargeable Nickel-MetalHydride (NiMH) battery.<br />

en<br />

WARNING<br />

Do not open the unit to replace the battery for any reason as there is the risk of a<br />

short-circuit. Opening the device will invalidate the warranty.<br />

The battery can be replaced only by an authorized service center.<br />

The battery should be fully charged before use.<br />

Recharging needs a certain period of time, while the unit normally operates when the micromotor<br />

is rotating.<br />

When connecting for charging, check that the green LED on the charger is flashing to show<br />

that the charger is properly connected to the power supply.<br />

The battery LED is lit in 3 colours indicating the current battery condition.<br />

Green: battery has sufficient capacity.