WARNING<br />
• Combined length determination by using the contra-angle and lip clip give only accurate<br />
results when using the original Sirona <strong>VDW</strong> 6:1 contra-angle.<br />
• Use only endodontic NiTi files with a metal handle for this type of length determination.<br />
• In some clinical cases accurate length determination is impossible! (For details see<br />
chapter 9 “Troubleshooting”)<br />
• Electric length determination uses minimal auxiliary current. The values used in the<br />
<strong>VDW</strong>.GOLD are far below the values required in IEC 601-1: 1988. Nevertheless, in rare<br />
cases electrical sensation during the length determination can occur. In this case, do not<br />
continue the treatment with this patient.<br />
4.9.1 Useful tips for accurate length determination<br />
• Gloves and rubber dam are re<strong>com</strong>mended in order to isolate the tooth.<br />
• Dry the access cavity with the suction pump or a cotton-pellet.<br />
• Avoid any direct contact between the contra-angle file and mucosa. Use the silicone<br />
sleeve for the contra-angle.<br />
4.9.2 Operation of Apex Locator<br />
For <strong>com</strong>bined length determination use the contra-angle, covered by the silicone sleeve.<br />
• Connect the lip clip cable to the mini USB (B) port on the back of the unit. Insert the lip<br />
clip into the connector at the end of the cable. Ensure the connection is solid.<br />
As soon as the mini USB connector is inserted into the port, the Apex Status LED (12)<br />
turns to red (when the unit is switched on, the LED is off). This means the unit has<br />
recognised the connector and is setting up to operate. Removing the lip clip cable from<br />
the mini USB port causes the LED to turn off again or switch off the device.<br />
• Put the lip clip in the patient´s mouth (we re<strong>com</strong>mend positioning it on the opposite side<br />
of the tooth undergoing therapy).<br />