03.03.2013 Views

Accutorr Plus Service Manual - Mindray

Accutorr Plus Service Manual - Mindray

Accutorr Plus Service Manual - Mindray

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

4.2 SAFETY PRECAUTIONS<br />

There are areas where improper repair techniques will cause further damage. When the<br />

instrument covers are removed, observe the following precautions and guidelines.<br />

1. Power line voltages will be exposed; identify hazard points and avoid direct contact.<br />

2. Battery terminals are accessible; do not short to each other or the positive terminal to<br />

ground circuits, as the capacity of the battery is sufficient to create an over heating<br />

safety hazard.<br />

3. Fuses must not be by passed or replaced by different current, voltage or time delay<br />

ratings.<br />

4. The circuit boards contain static sensitive components; use proper static safeguard<br />

techniques.<br />

CAUTION: Li-lon batteries are intended for replacement by qualified service personnel only.<br />

CAUTION: Li-lon batteries used in this device may present a risk of fire or chemical burn if<br />

mistreated. Do not disassemble, heat above 100C (212F), or incinerate. Replace battery with<br />

Datascope P/N: 0146-00-0069 only. Use of another battery may present a risk of fire or<br />

explosion.<br />

CAUTION: Dispose of used battery promptly. Keep away from children. Do not disassemble and<br />

do not dispose of in fire.<br />

4.3 GENERAL TROUBLESHOOTING GUIDELINES<br />

This procedure does not cover all possible problems. The intent of troubleshooting is to<br />

provide quick isolation and remedy of malfunctions and to return the instrument<br />

promptly to safe patient care. A logical approach will quickly identify the problem and<br />

suggest cost effective solutions.<br />

Troubleshooting Tips<br />

1. Isolate the cause of the problem to be either clinical, instrument or external.<br />

If a clinical cause is suspected, take the physiological measurement by a manual method<br />

or use an alternate measuring device. Check the physiological value against the instrument<br />

specifications; if exceeds specification, continue with manual or alternate method.<br />

If a problem with the instrument is suspected, first examine and / or replace patient<br />

connected accessories. Check instrument set-up / configuration. Ensure the correct<br />

battery charge profile is selected for the installed battery. If all set-ups are OK,<br />

replace instrument with similar device.<br />

If the environment is suspected, observe if the malfunction coincides with other<br />

equipment cycling (especially high power motors, RF generators, etc...). Replace<br />

problem instrument with an identical substitute. Place the problem instrument in a<br />

low disturbance environment. If the problem persists in the low disturbance<br />

environment, the instrument performance may be suspect. If the substitute instrument<br />

exhibits similar poor performance in the original environment, the environmental<br />

causes must be eliminated.<br />

4-2 Revised 12/20/00<br />

<strong>Accutorr</strong> <strong>Plus</strong> <strong>Service</strong> <strong>Manual</strong><br />

Chapter 4 - Repair Information

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!