1.3.10.3 Sequence for establishing SpO2 with Masimo Set Pulse Oximetry*<br />

* This feature applicable only if available or installed on your unit.<br />

1. Select the appropriate sensor for the patient from the table below. All sensors below are<br />

non-sterile and can be used during patient movement.<br />

MASIMO ® SENSOR FAMILY<br />

SELECTION PART NUMBER PATIENT SIZE DISPOSABLE/<br />

REUSABLE<br />

LNOP ® •Adt Adult<br />

Disposable Finger Sensor<br />

0600-00-0043-01 > 30 kg. Disposable<br />

LNOP ®<br />

•Pdt Pediatric/<br />

Slender Digit Disposable<br />

Sensor<br />

0600-00-0044-01 10 to 50 kg. Disposable<br />

LNOP ®<br />

•Neo Neonatal<br />

Disposable Sensor<br />

0600-00-0045-01 30 kg. Re-usable<br />

Reusable Finger Sensor<br />

PC12 Patient Cable<br />

Extension<br />

0012-00-1099-02 All Re-usable<br />

2. Attach the PC12 Patient Cable (P/N 0012-00-1099-02) to the sensor and plug the other<br />

end of the patient cable into the SpO2 connector (15)<br />

NOTE: Do not place the sensor on an extremity with an invasive catheter or blood pressure<br />

cuff in place.<br />

NOTE: Ensure proper routing of patient cable to avoid entanglement and/or<br />

strangulation.<br />

CAUTION: When equipped with MASIMO ® SpO2, use only MASIMO ® oxygen transducers<br />

including MASIMO LNOP ® patient dedicated adhesive sensors and MASIMO<br />

PC12 ® Patient Cable. Use of other oxygen transducers may cause improper oximeter<br />

performance.<br />

CAUTION: Tissue damage or inaccurate measurements may be caused by incorrect sensor<br />

application or use, such as wrapping it too tightly, applying supplemental tape,<br />

failing to inspect the sensor site periodically, or failing to position it appropriately.<br />

Carefully read the sensor directions for use, the <strong>Accutorr</strong> <strong>Plus</strong> operating instructions,<br />

and all precautionary information before use.<br />

CAUTION: Excessive ambient light may cause inaccurate measurements. Cover the sensor site<br />

with opaque material.<br />

CAUTION: Inaccurate measurements may be caused by incorrect sensor application or use;<br />

significant levels of dysfunctional hemoglobins, (e.g., carboxyhemoglobin or<br />

methemoglobin); or intra-vascular dyes such as indocyanine green methylene blue;<br />

exposure to excessive illumination, such as surgical lamps (especially ones with a<br />

xenon light source), bilirubin lamps, fluorescent lights, infrared heating lamps, or<br />

direct sunlight; excessive patient movement; venous pulsations; electro-surgical<br />

1-40 Revised 06/19/01<br />