1.3.10.3 Sequence for establishing SpO2 with Masimo Set Pulse Oximetry*<br />
* This feature applicable only if available or installed on your unit.<br />
1. Select the appropriate sensor for the patient from the table below. All sensors below are<br />
non-sterile and can be used during patient movement.<br />
MASIMO ® SENSOR FAMILY<br />
SELECTION PART NUMBER PATIENT SIZE DISPOSABLE/<br />
REUSABLE<br />
LNOP ® •Adt Adult<br />
Disposable Finger Sensor<br />
0600-00-0043-01 > 30 kg. Disposable<br />
LNOP ®<br />
•Pdt Pediatric/<br />
Slender Digit Disposable<br />
Sensor<br />
0600-00-0044-01 10 to 50 kg. Disposable<br />
LNOP ®<br />
•Neo Neonatal<br />
Disposable Sensor<br />
0600-00-0045-01 30 kg. Re-usable<br />
Reusable Finger Sensor<br />
PC12 Patient Cable<br />
Extension<br />
0012-00-1099-02 All Re-usable<br />
2. Attach the PC12 Patient Cable (P/N 0012-00-1099-02) to the sensor and plug the other<br />
end of the patient cable into the SpO2 connector (15)<br />
NOTE: Do not place the sensor on an extremity with an invasive catheter or blood pressure<br />
cuff in place.<br />
NOTE: Ensure proper routing of patient cable to avoid entanglement and/or<br />
strangulation.<br />
CAUTION: When equipped with MASIMO ® SpO2, use only MASIMO ® oxygen transducers<br />
including MASIMO LNOP ® patient dedicated adhesive sensors and MASIMO<br />
PC12 ® Patient Cable. Use of other oxygen transducers may cause improper oximeter<br />
performance.<br />
CAUTION: Tissue damage or inaccurate measurements may be caused by incorrect sensor<br />
application or use, such as wrapping it too tightly, applying supplemental tape,<br />
failing to inspect the sensor site periodically, or failing to position it appropriately.<br />
Carefully read the sensor directions for use, the <strong>Accutorr</strong> <strong>Plus</strong> operating instructions,<br />
and all precautionary information before use.<br />
CAUTION: Excessive ambient light may cause inaccurate measurements. Cover the sensor site<br />
with opaque material.<br />
CAUTION: Inaccurate measurements may be caused by incorrect sensor application or use;<br />
significant levels of dysfunctional hemoglobins, (e.g., carboxyhemoglobin or<br />
methemoglobin); or intra-vascular dyes such as indocyanine green methylene blue;<br />
exposure to excessive illumination, such as surgical lamps (especially ones with a<br />
xenon light source), bilirubin lamps, fluorescent lights, infrared heating lamps, or<br />
direct sunlight; excessive patient movement; venous pulsations; electro-surgical<br />
1-40 Revised 06/19/01<br />
<strong>Accutorr</strong> <strong>Plus</strong> <strong>Service</strong> <strong>Manual</strong><br />
Chapter 1, Operation