Alarm Limit Table<br />
PARAMETER RANGE UNITS FACTORY UNITS OF<br />
Systolic High<br />
DEFAULT INCREMENT<br />
Adult<br />
Off, 60-260<br />
Pediatric Off, 60-160 mmHg Off 5<br />
Neonate<br />
Systolic Low<br />
Off, 50-125<br />
Adult<br />
Off, 55-150<br />
Pediatric Off, 55-130 mmHg Off 5<br />
Neonate<br />
Diastolic High<br />
Off, 45-115<br />
Adult<br />
Off, 40-200<br />
Pediatric Off, 40-150 mmHg Off 5<br />
Neonate<br />
Diastolic Low<br />
Off, 35-100<br />
Adult<br />
Off, 30-120<br />
Pediatric Off, 30-50 mmHg Off 5<br />
Neonate<br />
MAP High<br />
Off, 25-50<br />
Adult<br />
Off, 90-200<br />
Pediatric Off, 90-150 mmHg Off 5<br />
Neonate<br />
MAP Low<br />
Off, 60-110<br />
Adult<br />
Off, 40-100<br />
Pediatric Off, 40-70 mmHg Off 5<br />
Neonate<br />
Pulse Rate High<br />
Off, 30-70<br />
Adult<br />
Off, 100-245<br />
Pediatric Off, 100-245 bpm Off 5<br />
Neonate<br />
Pulse Rate Low<br />
Off, 100-245<br />
Adult<br />
Off, 35-120<br />
Pediatric Off, 35-150 bpm Off 5<br />
Neonate<br />
SpO2 High<br />
Off, 75-200<br />
Adult<br />
Off, 61-99<br />
Pediatric Off, 61-99 %SpO2 Off 1<br />
Neonate<br />
SpO2 Low<br />
Off,61-99<br />
Adult<br />
60-95<br />
Pediatric 60-95<br />
%SpO2 86 1<br />
Neonate 60-95<br />
1.3.5.2 Alarm Violations<br />
An alarm condition exists if the parameter is equal to or is outside the high/low limit<br />
range that has been set. When an alarm limit is violated, the following actions occur:<br />
The LEDs for the parameter in an alarm condition flashes.<br />
The parameter in an alarm condition is in reverse video on the LCD (<strong>Accutorr</strong> <strong>Plus</strong> models<br />
with Trend Screen).<br />
The alarm tone is sounded (unless muted with the MUTE key (30)).<br />
The parameter(s) that was in an alarm condition will be in brackets [ ] when printed on<br />
the recorder.<br />
1-26 <strong>Accutorr</strong> <strong>Plus</strong> <strong>Service</strong> <strong>Manual</strong><br />