Alarm Limit Table<br />

PARAMETER RANGE UNITS FACTORY UNITS OF<br />

Systolic High<br />

DEFAULT INCREMENT<br />

Adult<br />

Off, 60-260<br />

Pediatric Off, 60-160 mmHg Off 5<br />

Neonate<br />

Systolic Low<br />

Off, 50-125<br />

Adult<br />

Off, 55-150<br />

Pediatric Off, 55-130 mmHg Off 5<br />

Neonate<br />

Diastolic High<br />

Off, 45-115<br />

Adult<br />

Off, 40-200<br />

Pediatric Off, 40-150 mmHg Off 5<br />

Neonate<br />

Diastolic Low<br />

Off, 35-100<br />

Adult<br />

Off, 30-120<br />

Pediatric Off, 30-50 mmHg Off 5<br />

Neonate<br />

MAP High<br />

Off, 25-50<br />

Adult<br />

Off, 90-200<br />

Pediatric Off, 90-150 mmHg Off 5<br />

Neonate<br />

MAP Low<br />

Off, 60-110<br />

Adult<br />

Off, 40-100<br />

Pediatric Off, 40-70 mmHg Off 5<br />

Neonate<br />

Pulse Rate High<br />

Off, 30-70<br />

Adult<br />

Off, 100-245<br />

Pediatric Off, 100-245 bpm Off 5<br />

Neonate<br />

Pulse Rate Low<br />

Off, 100-245<br />

Adult<br />

Off, 35-120<br />

Pediatric Off, 35-150 bpm Off 5<br />

Neonate<br />

SpO2 High<br />

Off, 75-200<br />

Adult<br />

Off, 61-99<br />

Pediatric Off, 61-99 %SpO2 Off 1<br />

Neonate<br />

SpO2 Low<br />

Off,61-99<br />

Adult<br />

60-95<br />

Pediatric 60-95<br />

%SpO2 86 1<br />

Neonate 60-95<br />

1.3.5.2 Alarm Violations<br />

An alarm condition exists if the parameter is equal to or is outside the high/low limit<br />

range that has been set. When an alarm limit is violated, the following actions occur:<br />

The LEDs for the parameter in an alarm condition flashes.<br />

The parameter in an alarm condition is in reverse video on the LCD (<strong>Accutorr</strong> <strong>Plus</strong> models<br />

with Trend Screen).<br />

The alarm tone is sounded (unless muted with the MUTE key (30)).<br />

The parameter(s) that was in an alarm condition will be in brackets [ ] when printed on<br />

the recorder.<br />