If patient movement presents a problem:<br />
Verify that the sensor is properly and securely applied.<br />
Use a new sensor with fresh adhesive backing.<br />
Move the sensor to a less active site.<br />
Use a type of sensor that tolerates some patient motion, such as the OXISENSOR<br />
D-25, D-20, N-25, or I20 oxygen transducer.<br />
NELLCOR® SENSOR FAMILY<br />
SELECTION D25/D25L R-15 N-25 I-20 D-20 RS-10<br />
GUIDE Adult Adult Neonatal Infant Pediatric Adult<br />
Patient >30 kg >50 kg 40 kg<br />
Size<br />
>40 kg<br />
Duration Short or Long Short or Short or Short or Short or Short Term<br />
of Use Term Long Term Long Term Long Term Long Term<br />
Sterility Sterile1 Sterile1 Sterile1 Sterile1 Sterile1 Nonsterile<br />
Patient Limited Inactive Limited Limited Limited Limited<br />
Activity Activity<br />
Activity Activity Activity Activity<br />
OXISENSOR OXISENSOR OXISENSOR OXISENSOR OXISENSOR RS-10<br />
adult digit adult nasal neonatal infant digit pediatric reflectance<br />
oxygen oxygen oxygen oxygen digit oxygen oxygen<br />
transducer transducer transducer transducer transducer transducer<br />
1 In an unopened, undamaged package.<br />
All NELLCOR ® accessories and sensors must be purchased form NELLCOR ® Inc.,<br />
25495 Whitehall Street, Hayward, Ca. 94545. To contact NELLCOR ® , call<br />
1-800-NELLCOR.<br />
D. Automatic Calibration of NELLCOR ® Sensors<br />
The oximetry subsystem incorporates automatic calibration mechanisms. It is<br />
automatically calibrated each time it is turned on, at periodic intervals thereafter, and<br />
whenever a new sensor is connected. Also, the intensity of the sensor’s LEDs is<br />
adjusted automatically to compensate for differences in tissue thickness.<br />
Each sensor is calibrated when manufactured; the effective mean wavelength of the red<br />
LED is determined and encoded into a calibration resistor in the sensor plug. The<br />
instrument’s software reads this calibration resistor to determine the appropriate<br />
calibration coefficients for the measurements obtained by that sensor.<br />
<strong>Accutorr</strong> <strong>Plus</strong> <strong>Service</strong> <strong>Manual</strong> Revised 02/15/00<br />
1-39<br />
Chapter 1, Operation