If patient movement presents a problem:<br />

Verify that the sensor is properly and securely applied.<br />

Use a new sensor with fresh adhesive backing.<br />

Move the sensor to a less active site.<br />

Use a type of sensor that tolerates some patient motion, such as the OXISENSOR<br />

D-25, D-20, N-25, or I20 oxygen transducer.<br />

NELLCOR® SENSOR FAMILY<br />

SELECTION D25/D25L R-15 N-25 I-20 D-20 RS-10<br />

GUIDE Adult Adult Neonatal Infant Pediatric Adult<br />

Patient >30 kg >50 kg 40 kg<br />

Size<br />

>40 kg<br />

Duration Short or Long Short or Short or Short or Short or Short Term<br />

of Use Term Long Term Long Term Long Term Long Term<br />

Sterility Sterile1 Sterile1 Sterile1 Sterile1 Sterile1 Nonsterile<br />

Patient Limited Inactive Limited Limited Limited Limited<br />

Activity Activity<br />

Activity Activity Activity Activity<br />

OXISENSOR OXISENSOR OXISENSOR OXISENSOR OXISENSOR RS-10<br />

adult digit adult nasal neonatal infant digit pediatric reflectance<br />

oxygen oxygen oxygen oxygen digit oxygen oxygen<br />

transducer transducer transducer transducer transducer transducer<br />

1 In an unopened, undamaged package.<br />

All NELLCOR ® accessories and sensors must be purchased form NELLCOR ® Inc.,<br />

25495 Whitehall Street, Hayward, Ca. 94545. To contact NELLCOR ® , call<br />

1-800-NELLCOR.<br />

D. Automatic Calibration of NELLCOR ® Sensors<br />

The oximetry subsystem incorporates automatic calibration mechanisms. It is<br />

automatically calibrated each time it is turned on, at periodic intervals thereafter, and<br />

whenever a new sensor is connected. Also, the intensity of the sensor’s LEDs is<br />

adjusted automatically to compensate for differences in tissue thickness.<br />

Each sensor is calibrated when manufactured; the effective mean wavelength of the red<br />

LED is determined and encoded into a calibration resistor in the sensor plug. The<br />

instrument’s software reads this calibration resistor to determine the appropriate<br />

calibration coefficients for the measurements obtained by that sensor.<br />

<strong>Accutorr</strong> <strong>Plus</strong> <strong>Service</strong> <strong>Manual</strong> Revised 02/15/00<br />

1-39<br />

Chapter 1, Operation