2. Use the proper equipment. Special test equipment required for problem isolation and<br />
resolution is listed in section 4.4. Similar devices may be used if they equal or exceed<br />
performance standards of the equipment suggested.<br />
PRECAUTION: This instrument utilizes multiple layer circuit boards and laser and<br />
ultra sonic welded surface mounted components. Do not attempt to un-solder welded<br />
components from the circuit board; the circuit board will be permanently damaged.<br />
PRECAUTION: Soldering and solder removal equipment must be low voltage operated<br />
and grounded to avoid static charge and stray current induced component damage.<br />
Maximum wattage: 25 W.<br />
3. Clean the repair area. After soldering operations, clean off the repaired area with ethyl<br />
or methyl alcohol and a stiff hair brush. This will remove residual solder flux, making<br />
the repaired area more visible for an inspection and returning the instrument to its<br />
original, neat appearance. Removal of the flux will also facilitate making electrical<br />
measurements in the affected area as the flux itself is not conductive.<br />
Exchange Program<br />
Datascope offers a comprehensive circuit board and electro-mechanical module exchange<br />
program. The exchange circuits and modules are warranted, factory pre-tested and<br />
calibrated. Final calibration of the exchange item is strongly suggested to match the new<br />
part to the host system. See section 6.4 for details concerning the exchange program.<br />
4.4 TEST EQUIPMENT REQUIRED<br />
Equipment types other than these listed may be utilized if they equal or exceed the listed<br />
equipment performance.<br />
• Dynatech Nevada Non-invasive Blood Pressure Simulator (Cuff Link)<br />
• Power Supply, Power-Mate, 3A 20V<br />
• Fluke D.V.M, Model 8050A<br />
• Digital Storage Scope, Tektronix, 2230 100 MHZ<br />
• Stop Watch, Wilson LW#119<br />
• SpO2 Simulator, Biotek Index, Smart SAT (All SpO2 versions);<br />
N-1290 (Nellcor SpO2 Only)<br />
• RS232 Test Connector (Pins 2 and 3 Shorted)<br />
• 700cc Dummy Cuff, Datascope #0138-00-0001-01<br />
• Chart Paper, Datascope #0683-00-0447-01<br />
• Water Bath, Cole-Palmer model H-12-105-10<br />
• Reference Quartz Thermometer, Hewlet-Packard model HP-2804A<br />
with Temp. Probe HP-1811A<br />
<strong>Accutorr</strong> <strong>Plus</strong> <strong>Service</strong> <strong>Manual</strong> Revised 12/20/00<br />
4-3<br />
Chapter 4 - Repair Information