7.6.2 Water Bath Method<br />

1. Prepare a water bath set to 89.6 to 110.0 F as a temperature reference. Confirm<br />

temperature with a separate precision thermometer.<br />

2. Snap a protective sheath onto the temperature probe, and immerse into the reference<br />

water bath for 30 seconds, or until the temperature display stops flashing. Record the<br />

temperature shown on the temperature display.<br />

3. Apply the correction factor, shown in table 7-13, to obtain the calibrated measurement.<br />

Table 7-13<br />

Celsius Readouts Fahrenheit Readouts<br />

Range Subtract Range Subtract<br />

32.0 to 41.0 0.6 89.6 to 102.0 1.1<br />

41.0 to 43.0 0.5 102.1 to 106.0 1.0<br />

106.0 to 110.0 .5<br />

4. Compare readouts after correction factors are calculated. Proceed to System Calibration<br />

Procedure if readings are out of specification.<br />

7.6.3 System Calibration Procedure<br />

1. Connect a fresh 9 volt battery or an external power supply set to 9 .1V to the<br />

predictive thermometer. NOTE: Newer modules do not contain a 9V battery. Skip<br />

this step.<br />

2. Set SW2-3 to closed position (the switch is towards the PCB). Connect resistor decade<br />

box to the temperature probe connector. Set the decade box to 8343 ohms, 2 ohms.<br />

NOTE: (All other settings of SW2 should be in the open position; away from the PCB.)<br />

3. Connect the DVM to J4-2 (-) and J4-7 (+) output jacks.<br />

4. Jumper J4-2 and ground terminal ( J2-1) jacks.<br />

5. Switch power supply on.<br />

6. Adjust R1 until a reading of 100 mV, 5mVisontheDVM.<br />

7. Adjust the decade box to 4746 ohms, 2 ohms. Record the voltage displayed on the<br />

DVM.<br />

8. Connect the DVM + lead to J4-4. Adjust R2 until the DVM reading is the same as<br />

the one recorded in step “7”, 1mV.<br />

9. Remove all jumpers and test leads. The calibration is completed.<br />

10. Set the decade box back to 8343 ohms. Power cycle the unit (switch off then on).<br />

The initial start-up temperature of 85.0 o F should be displayed.<br />

<strong>Accutorr</strong> <strong>Plus</strong> <strong>Service</strong> <strong>Manual</strong> Revised 01/11/00<br />

7-15<br />

Chapter 7 - Calibration