Table 4-1<br />
Error Codes<br />
4.5 TROUBLESHOOTING (PROBLEM ISOLATION)<br />
This section has been divided into two major areas; clinical and technical. Clinical problems<br />
can generally be resolved by the user. Technical problems may require resolution by a<br />
qualified technical service person. The error code table below indicates Technical problems<br />
with an asterisk (*). Error codes are displayed on the front panel LED’s of the <strong>Accutorr</strong> <strong>Plus</strong>.<br />
TYPE CODE DESCRIPTION REASON<br />
NIBP 8810 Retry - Unable to Motion artifact, cycle time-out, weak pulsations or no pulsations.<br />
Measure<br />
A triple beep tone is generated.<br />
8811 Retry - Pump<br />
Higher<br />
Insufficient cuff pressure. A triple beep tone is generated.<br />
8812 Stop - Cuff<br />
Excessive cuff pressure detected by the software. A triple beep<br />
Overpressure tone is generated.<br />
8813 Stop - Unable to 4 successive measurement attempts failed. A triple beep tone is<br />
Measure<br />
generated.<br />
TEMP 8830 Check Probe Tissue contact may have been lost.<br />
(PTM) 8831 Replace Probe Defective probe or connection.<br />
8832 Battery Low The 9V battery needs replacement.<br />
SpO2 8850 No Sensor No sensor connected.<br />
8851 Sensor Off Sensor not on patient.<br />
(Datascope and Masimo SpO2 only)<br />
8852 Interference Interference on signal.<br />
(Datascope and Masimo SpO2 only)<br />
8853 Pulse Search Unit cannot find signal. (Nellcor SpO2 Module will report<br />
“Pulse Search” -8853- when the sensor is not on the patient.)<br />
8854 Weak Pulse Weak pulse detected.<br />
(Datascope and Masimo SpO2 only)<br />
8855 No Pulse No pulse detected. (Datascope SpO2 only)<br />
8856 Check Sensor Sensor problem.<br />
(Datascope and Masimo SpO2 only)<br />
8857 PR250 Pulse rate is greater than 250 bpm. (Datascope SpO2 only)<br />
SYSTEM 984* NIBP Hardware Failure NIBP A/D failure detected.<br />
985* NIBP Overpressure<br />
Circuit not<br />
Programmed<br />
The overpressure circuit is not set to the current patient size.<br />
986* NIBP Overpressure<br />
Circuit not Tracking<br />
The two pressure transducers are not tracking each other.<br />
987* Stop - Hardware Excessive cuff pressure detected by hardware over-pressure<br />
Overpressure sensor. A triple beep tone is generated.<br />
988* TEMP Bad Calibration Thermometer needs calibration.<br />
990* TEMP Illegal Mode Thermometer switch is set wrong.<br />
991* TEMP Module Failed Thermometer internal failure.<br />
995* SpO2 Uncalibrated SpO2 fails calibration check.<br />
996* SpO2 Failure SpO2 failed self-test.<br />