Table 4-1<br />

Error Codes<br />

4.5 TROUBLESHOOTING (PROBLEM ISOLATION)<br />

This section has been divided into two major areas; clinical and technical. Clinical problems<br />

can generally be resolved by the user. Technical problems may require resolution by a<br />

qualified technical service person. The error code table below indicates Technical problems<br />

with an asterisk (*). Error codes are displayed on the front panel LED’s of the <strong>Accutorr</strong> <strong>Plus</strong>.<br />

TYPE CODE DESCRIPTION REASON<br />

NIBP 8810 Retry - Unable to Motion artifact, cycle time-out, weak pulsations or no pulsations.<br />

Measure<br />

A triple beep tone is generated.<br />

8811 Retry - Pump<br />

Higher<br />

Insufficient cuff pressure. A triple beep tone is generated.<br />

8812 Stop - Cuff<br />

Excessive cuff pressure detected by the software. A triple beep<br />

Overpressure tone is generated.<br />

8813 Stop - Unable to 4 successive measurement attempts failed. A triple beep tone is<br />

Measure<br />

generated.<br />

TEMP 8830 Check Probe Tissue contact may have been lost.<br />

(PTM) 8831 Replace Probe Defective probe or connection.<br />

8832 Battery Low The 9V battery needs replacement.<br />

SpO2 8850 No Sensor No sensor connected.<br />

8851 Sensor Off Sensor not on patient.<br />

(Datascope and Masimo SpO2 only)<br />

8852 Interference Interference on signal.<br />

(Datascope and Masimo SpO2 only)<br />

8853 Pulse Search Unit cannot find signal. (Nellcor SpO2 Module will report<br />

“Pulse Search” -8853- when the sensor is not on the patient.)<br />

8854 Weak Pulse Weak pulse detected.<br />

(Datascope and Masimo SpO2 only)<br />

8855 No Pulse No pulse detected. (Datascope SpO2 only)<br />

8856 Check Sensor Sensor problem.<br />

(Datascope and Masimo SpO2 only)<br />

8857 PR250 Pulse rate is greater than 250 bpm. (Datascope SpO2 only)<br />

SYSTEM 984* NIBP Hardware Failure NIBP A/D failure detected.<br />

985* NIBP Overpressure<br />

Circuit not<br />

Programmed<br />

The overpressure circuit is not set to the current patient size.<br />

986* NIBP Overpressure<br />

Circuit not Tracking<br />

The two pressure transducers are not tracking each other.<br />

987* Stop - Hardware Excessive cuff pressure detected by hardware over-pressure<br />

Overpressure sensor. A triple beep tone is generated.<br />

988* TEMP Bad Calibration Thermometer needs calibration.<br />

990* TEMP Illegal Mode Thermometer switch is set wrong.<br />

991* TEMP Module Failed Thermometer internal failure.<br />

995* SpO2 Uncalibrated SpO2 fails calibration check.<br />

996* SpO2 Failure SpO2 failed self-test.<br />

4-4 Revised 02/15/00<br />

<strong>Accutorr</strong> <strong>Plus</strong> <strong>Service</strong> <strong>Manual</strong><br />

Chapter 4 - Repair Information