3.0 SPECIFICATIONS<br />

CONTENTS OF THIS CHAPTER ...........................Page<br />

3.1 Performance Specifications ..............................3-1<br />

3.2 Safety Characteristics ..................................3-6<br />

3.3 Physical Characteristics ................................3-7<br />

3.4 Environmental Characteristics ...........................3-7<br />

3.5 Electrical Ratings .....................................3-7<br />

3.6 Agency Compliance ...................................3-8<br />

3.7 Electromagnetic Compatibility...........................3-8<br />

3.1 PERFORMANCE SPECIFICATIONS<br />

Systolic Pressure Readout<br />

Number of Digits: 3<br />

Accuracy*: Mean error less than 5 mmHg, Standard deviation<br />

less than 8 mmHg.<br />

Range: Adult Mode: 55 to 260 mmHg<br />

Pediatric Mode: 55 to 160 mmHg<br />

Neonatal Mode: 45to 120 mmHg<br />

Diastolic Pressure Readout<br />

Number of Digits: 3<br />

Accuracy*: Mean error less than 5 mmHg, Standard deviation<br />

less than 8 mmHg.<br />

Range: Adult Mode: 30 to 200 mmHg<br />

Pediatric Mode: 30 to 150 mmHg<br />

Neonatal Mode: 25 to 100 mmHg<br />

*Tested per ANSI/AAMI SP10-1992 methods.<br />

<strong>Accutorr</strong> <strong>Plus</strong> <strong>Service</strong> <strong>Manual</strong> 3-1<br />

Revised 12/20/00<br />

Chapter 3 - Specifications