<strong>SR715</strong>/720 <strong>LCR</strong> METERTypical measurement setupsThe table below lists suggested test conditions for various types of components. This is by no means the onlyset of conditions for measuring, but is a good general "rule of thumb".Setups for Measuring Typical PartsComponentTypeValueParameterTypeEquivalentCircuitFrequencyUnknown: Any Auto series 1 kHzResistors: < 1 kΩ R+Q series 1 kHz> 1 kΩ R+Q series 100 or 120 Hz< 10 µH L+Q series 100 kHz10 µH – 1 mH L+Q series 10 kHz1 mH – 1 H L+Q series 1 kHz> 1 H L+Q series 100 or 120 Hz 1 µF C+R or C+D series 100 or 120 Hzx