Stratos®Pro A4... PH User Manual
Stratos®Pro A4... PH User Manual
Stratos®Pro A4... PH User Manual
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Step Action/Display Remark<br />
Select HOLD<br />
mode<br />
Disconnect old<br />
sensor,<br />
connect new<br />
sensor.<br />
Wait until the<br />
sensor data are<br />
displayed.<br />
Check sensor<br />
data<br />
Check<br />
measured<br />
values,<br />
then<br />
exit HOLD.<br />
Press menu key to call<br />
the selection menu,<br />
select HOLD using the<br />
��keys, press enter<br />
to confirm.<br />
MEMO SENS<br />
MEMO SENS<br />
View sensor information<br />
using �� keys,<br />
press enter to confirm.<br />
Hit meas key:<br />
Return to the selection<br />
menu. Hold meas<br />
key depressed: Device<br />
switches to measuring<br />
mode.<br />
Digital Sensors<br />
Now the device is in HOLD<br />
mode. The HOLD mode can also<br />
be activated externally via the<br />
HOLD input. During HOLD the<br />
output current is frozen at its last<br />
value or set to a fixed value.<br />
Temporary messages are displayed<br />
during the replacement<br />
but neither output to the alarm<br />
contact nor entered in the logbook.<br />
You can view the sensor manufacturer<br />
and type, serial number,<br />
and last calibration date.<br />
The sensor replacement is<br />
entered in the extended<br />
logbook.<br />
121