Step Action/Display Remark<br />

Select HOLD<br />

mode<br />

Disconnect old<br />

sensor,<br />

connect new<br />

sensor.<br />

Wait until the<br />

sensor data are<br />

displayed.<br />

Check sensor<br />

data<br />

Check<br />

measured<br />

values,<br />

then<br />

exit HOLD.<br />

Press menu key to call<br />

the selection menu,<br />

select HOLD using the<br />

��keys, press enter<br />

to confirm.<br />

MEMO SENS<br />

MEMO SENS<br />

View sensor information<br />

using �� keys,<br />

press enter to confirm.<br />

Hit meas key:<br />

Return to the selection<br />

menu. Hold meas<br />

key depressed: Device<br />

switches to measuring<br />

mode.<br />

Digital Sensors<br />

Now the device is in HOLD<br />

mode. The HOLD mode can also<br />

be activated externally via the<br />

HOLD input. During HOLD the<br />

output current is frozen at its last<br />

value or set to a fixed value.<br />

Temporary messages are displayed<br />

during the replacement<br />

but neither output to the alarm<br />

contact nor entered in the logbook.<br />

You can view the sensor manufacturer<br />

and type, serial number,<br />

and last calibration date.<br />

The sensor replacement is<br />

entered in the extended<br />

logbook.<br />